Preview

Radiatsionnaya Gygiena = Radiation Hygiene

Advanced search

Evaluation of levels of exposure of adult patients from common radiographic examinations in the Russian Federation in 2009–2014

https://doi.org/10.21514/1998-426X-2017-10-3-66-75

Abstract

Diagnostic reference levels are the main and the most effective tools of optimization of the radiation protection of patients from medical exposure. Diagnostic reference levels should be established based on the results of dedicated dose surveys, allowing evaluating typical patient dose distributions in a selected dose quantity for the selected X-ray examinations. The aim of the current study was to assess the distributions of typical effective doses in representative Russian regions. Materials and methods: Typical patient effective doses for the 13 most common radiographic X-ray examinations were collected in 203 X-ray rooms in 101 hospitals in six regions of Russian Federation in 2009–2014. A differentiated approach was used for the estimation of the typical effective doses depending on the image acquisition technology. Effective doses were estimated using «EDEREX» (Russia) computational software. Results and discussion: Results of the dose data analysis indicate the lack of significant differences between the distributions of the typical effective doses between the selected regions, allowing merging the regional samples and further evaluating the pooled (joint) sample. A significant ratio of maximum to minimum (up to two orders of magnitude) due to a presence of X-ray units with abnormally high and low typical effective doses was observed for all 13 selected X-ray examinations. Abnormally high typical effective doses can be explained by performing the examinations using high values of tube current-time product (150–600 mAs) on a maximum field size (up 40×40 cm). Removal of the typical effective doses below 5%-percentile and above 95%-percentile of typical effective dose distributions for all examinations would result in a reduction of a mean effective dose by up to 30% and reduction of a 75%-percentile of the distributions by up to 15%. No significant differences between the distributions of TED for analogue and digital X-ray units were observed for the pooled sample for selected examinations except for the examination of the chest in posterior-anterior projection. Conclusions: These results should be considered in the process of establishing and implementing DRLs as well as in the cost-benefit analysis of the optimization in radiography. 

About the Authors

A. V. Vodovatov
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation

Aleksandr V. Vodovatov – Research scientist 

For correspondence: Mira Str., 8, St. Petersburg, 197101, Russia; E-mail: vodovatoff@gmail.com



V. Yu. Golikov
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation
Vladislav Yu.Golikov – Senior Researcher


S. A. Kalnitsky
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation
Sergey A. Kalnitsky – Leading Researcher


I. G. Shatsky
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation
Ilya G.Shatsky – Researcher


L. A. Chipiga
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation
Larisa A. Chipiga – Engineer researcher


References

1. Vishnyakova N.M. [et al.] Methodical aspects of establishment of diagnostic reference levels for radiography for adult patients. Vestnik Rossiyskoy Voenno-meditsinskoy akademii = News of Russian Military Medical Academy, 2010, № 1 (29), pp. 96–102. (In Russian).

2. Vodovatov A. V. Practical implementation of diagnostic reference levels (DRLs) concept for the common radiographic examinations. Radiatsionnaya Gygiena = Radiation Hygiene, 2017, Vol. 10, № 1, pp. 47-55. (In Russian).

3. ICRP, 201x. Diagnostic Reference Levels in Medical Imaging. ICRP Publication 1XX Ann. ICRP 4X(X-X). – Available on: http://www.icrp.org/ docs/C3WPDRLDraftForPublicConsult ation(011116).pdf (Accessed: 04.08.2017).

4. Repin V.S. [et al.] Doses to the public of Russian Federation based on the ESKID results in 2002-2015. Informatsionnyy sbornik = Information bulletin. St-Petersburg, 2015, 40 p. (In Russian).

5. Vodovatov A.V., Balonov M.I., Golikov V.Yu. [et. al.] Proposals for the establishment of national diagnostic reference levels for radiography for adult patients based on regional dose surveys in Russian Federation. Rad. Prot. Dosim, 2017, Vol. 173, № 1-3, pp. 223-232.

6. Golikov V.Yu. [et al.] Assessment of effective doses of patients from diagnostic X-ray examinations. Proceedings «Radiation Hygiene». St-Petersburg, 2003, pp. 75–88. (In Russian).

7. Petrie A., Sabin C. Medical Statistics at a Glance. Moscow, GEOTAR-Media, 2010, 168 p. (In Russian).

8. Vodovatov A.V., Kamyshanskaya I.G., Drozdov A.A. New approach for the determination of the standard patient to be used for optimization of the medical exposure protection. Radiatsionnaya Gygiena = Radiation Hygiene, 2014, Vol.7, No 4, pp. 104-116. (In Russian).


Review

For citations:


Vodovatov A.V., Golikov V.Yu., Kalnitsky S.A., Shatsky I.G., Chipiga L.A. Evaluation of levels of exposure of adult patients from common radiographic examinations in the Russian Federation in 2009–2014. Radiatsionnaya Gygiena = Radiation Hygiene. 2017;10(3):66-75. (In Russ.) https://doi.org/10.21514/1998-426X-2017-10-3-66-75