Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM

https://doi.org/10.21514/1998-426X-2019-12-4-89-95

Abstract

This work is devoted to the assessment of the absorbed dose in the skin of patients undergoing interventional radiological examinations. There is a probability of deterministic effects in the skin of patients due to high doses of radiation during this type of medical examinations. The aim of this work was to conduct direct measurements of the absorbed dose in the skin of patients undergoing interventional radiological procedures using special dosimetric radiochromic films Gafchromic XR-RV3 to visualize the distribution of radiation over the patient’s skin surface and to study the possibility of exceeding the threshold values for the deterministic effects in the skin. The paper discusses the features of measurements with GafchromiC XR-RV3 films. The method of film digitization on conventional flatbed scanner and image processing with the ImageJ program was tested. The obtained calibration curve for this type of film was presented. The skin dose distributions for several interventional studies obtained with radiochromic films were presented. The measured value of the maximum absorbed dose in the skin for four of the fourteen analyzed procedures exceeded the threshold value of the absorbed dose for the occurrence of skin erythema in 2 Gy. The highest values of the maximum absorbed dose in the skin were obtained for the coronary angioplasty – 3.2 Gy and for the procedure of embolization of the uterine arteries – 2.9 Gy.

About the Author

Svetlana S. Sarycheva
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev, Federal Service for Surveillance on Consumer Rights Protection and Human Well-Being
Russian Federation

candidate of biological sciences, Senior Scientific Researcher

Mira Str., 8, St. Petersburg, 197101



References

1. UNSCEAR UN. Medical exposure to ionizing radiation. United National Scientific Committee on the Effects of Atomic Radiation. A/AC.82/R. 669. Vienna: United Nations, 2008, 301 p.

2. ICRP. Publication 85. Avoidance of radiation injuries from medical interventional procedures. Ann. ICRP. Vienna, Pergamon Press, 2000, vol. 30, № 2, 68 p.

3. Wagner L.K. [et al.] Severe skin reactions from interventional fluoroscopy: Case report and review of literature. Radiology 1999; 213: pp. 773–776.

4. Galstyan I.A., Nadezhina N.M. Local Radiation Injuries as Complications of Medical Radiation Exposure. Meditsinskaya radiologiya i radiatsionnaya bezopasnost = Medical Radiology and Radiation Safety, 2012, 57 (5), pp. 31-36. (in Russian)

5. Amiraslanov Y.А. Local radiation injuries gained due to endovascular interventions of coronary arteries / Y.А. Amiraslanov [et al.] // Diagnostic and Interventional Radiology, 2007, 2 (1), pp. 48-54. (in Russian)

6. Golikov V.Yu. Estimation of patient exposure under intervention radiological examinations / V.Yu. Golikov [et al.] // Radiatsionnaya Gygiena = Radiation Hygiene, SPb. – 2009. vol. 2, № 3, pp. 26-31. (in Russian)

7. Balter S. [et al.] Techniques to estimate radiation dose to skin during fluoroscopically guided procedures. J. Vasc. Interv. Radiol., 2001, Vol. 13, pp. 391-397.

8. Miller D.L. [et al.] Radiation doses in interventional radiology procedures: The RAD-IR study: Part II: Skin Dose. J. Vasc. Interv. Radiol., 2003, Vol. 14, рр. 977-990.

9. Farah J. [et al.] Characterization of XR-RV3 GafChromic(®) films in standard laboratory and in clinical conditions and means to evaluate uncertainties and reduce errors. Med. Phys., 2015 Jul, 42(7), pp. 4211-4226.

10. Official site of the manufacturer of X-ray sensitive films Gafchromic. – Available on: http://www.gafchromic.com/gafchromic-film/radiology-films/XR-RV3/index.asp (Accessed: 20.04.2019) (In Russian)

11. Konyukhov A.L. Guidelines for using of ImageJ software for image processing: Training guide. Tomsk: department TU, TUSUR, 2012, 105 p. (in Russian)

12. McCabe B.P., Speidel M.A. Calibration of GafChromic XRRV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner. Med. Phys., 2011 Apr, 38(4), pp. 1919-1930.

13. ICRU. Publication 74. Patient Dosimetry for X-rays used in medical imaging. Journal of the ICRU, 2005, Vol 5, No 2, 116 p.

14. Rampado O. and R. Ropolo. Entrances skin dose distribution maps for interventional neuroradiological procedures: a preliminary study. Radiation protection dosimetry, 2005, Vol. 117, No 1-3, pp. 256-259.

15. Neocleous A. [et al.] Dosimetry using Gafchromic XR-RV2 radiochromic films in interventional radiology. Radiation Protection Dosimetry, September 2011, Vol. 147, issue 1-2, pр. 78-82.

16. Morell R.-E. Dosimetry and optimisation in high dose fluoroscopic and fluorographic procedures. University of Nottingham, March 2006, 215 p.

17. Rehani M. M., Srimahachota S. Skin injuries in interventional procedures. Radiat. Prot. Dosimetry, 2011, 147 (1-2), pp. 8-12.

18. Struelens L. [et al.] Skin dose measurements on patients for diagnostic and interventional neuroradiology: a multicentre study. Radiation protection dosimetry, 2005, Vol. 114, No. 1-3, pp. 143-146.


Review

For citations:


Sarycheva S.S. Patients skin dose measurements during interventional radiological examinations using Gafchromic XR-RV3 FILM. Radiatsionnaya Gygiena = Radiation Hygiene. 2019;12(4):89-95. (In Russ.) https://doi.org/10.21514/1998-426X-2019-12-4-89-95

Views: 1198


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1998-426X (Print)
ISSN 2409-9082 (Online)