Estimation of effective dose for children in interventional cardiology
https://doi.org/10.21514/1998-426X-2017-10-2-16-22
Abstract
This study is devoted to the estimation of effective dose for children undergoing interventional cardiology examinations. The conversion coefficients (CC) from directly measured dose area product (DAP) value to effective dose (ED) were calculated within the approved effective dose assessment methodology (Guidelines 2.6.1. 2944-11). The CC, Ed K , [mSv / (Gy • cm2)] for newborn infants and children of 1, 5, 10 and 15 years old (main(range)) were calculated as 2.5 (1.8-3.2); 1.1 (0.8-1.3); 0.6 (0.4-0.7); 0.4 (0.3-0.5); and 0,22 (0,18-0,30) respectively. A special Finnish computer program PCXMC 2.0 was used for calculating the dose CC. The series of calculations were made for different values of the physical and geometrical parameters based on their real-existing range of values. The value of CC from DAP to ED were calculated for all pediatric age groups. This work included 153 pediatric interventional studies carried out in two hospitals of the city of St. Petersburg for the period of one year from the summer of 2015. The dose CC dependency from the patient’s age and parameters of the examinations were under the study. The dependence from the beam quality (filtration and tube voltage) and age of the patient were found. The younger is the patient, stronger is the filtration and higher is the voltage, the higher is the CC value. The CC in the younger (newborn) and older (15 years) age groups are different by the factor of 10. It was shown that the changes of the geometric parameters (in the scope of their real existing range) have small effect on the value of the effective dose, not exceed 30-50% allowable for radiation protection purpose. The real values of effective doses of children undergoing cardiac interventions were estimated. In severe cases, the values of ED can reach several tens of mSv.
About the Author
S. S. SarychevaRussian Federation
Svetlana S. Sarycheva - Senior Scientific Researcher
Address for correspondence: Mira Str., 8, Saint-Petersburg, 197101, Russia; E-mail: Svetlana2003@mail.ru
References
1. Golikov V., Sarycheva S., Balonov M., Kalnitsky S. The patient dose assessment during the interventional radiological examinations. Radiatsionnaya Gygiena = Radiation Hygiene, 2009, Vol. 2, №3, pp. 75-88. (In Russian)
2. Sarycheva S., Golikov V., Kalnitsky S. Studies of patient doses in interventional radiological examinations. Radiation Protection Dosimetry, 2010, Vol. 139, No 1-3, pp. 258-261.
3. ICRP, 1991. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann. ICRP 21 (1-3).
4. ICRP, 1996. Conversion Coefficients for use in Radiological Protection against External Radiation. ICRP Publication 74. Ann. ICRP 26 (3-4).
5. Golikov V., Barkovsky A.N., Baryshkov N.K., Vlasov A.Yu. Evaluation of the effective doses to patients during radiological examinations. Radiatsionnaya Gygiena = Radiation Hygiene, SPb., 2003, pp. 75 – 88. (In Russian)
6. Tapiovaara M. A. Monte Carlo program for calculating patient doses in medical x-ray examinations. Ed.: M. Tapiovaara, T. Siiskonen. STUK-A231. November 2008, 49 p.
7. Rassow J., Schmaltz A.A., Hentrich F., Steffer C. Effective doses to patients from pediatric cardiac catheterization. Br J Radiol, , Vol. 73, pp. 172-183.
8. Karambatsakidou A. [et al.] Effective dose conversion factors in pediatric interventional cardiology. Br J Radiol, 2009, Vol. , pp. 748-755.
9. Axelsson B. [et al.] Estimating the effective dose to children undergoing heart investigations – a phantom study. Br J Radiol, , Vol. 72, pp. 378-383.
10. Schmidt P.W.E. Conversion factors for the estimation of effective dose in peadiatric cardiac angiography. Phys. Med. Biol., , Vol. 85, pp. 3095-3107.
11. Bacher K. [et al.] Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation, 2005, Vol. 111, pp. 83-89.
12. Schultz F.W., Geleijns J., Spoelstra F.M., Zoete J. Monte Carlo calculations for assessment of radiation dose to patients with congenital heart defect and to staff during cardiac catheterizations. Br J Radiol, 2003, Vol. 76, pp. 638-647.
13. Onnasch D.G.W., Schroder F.K., Fischer G., Kramer H.H. Diagnostic reference levels and effective dose in pediatric cardiac catheterization. Br J Radiol, 2007, Vol. 80, pp. 177-185.
14. Miksys N., Gordon C.L., Thomas K., Connolly B.L. Estimating effective dose to pediatric patients undergoing interventional radiology procedures using anthropomorphic phantoms and MOSFET dosimeters. AJR, 2010, Vol. 194, pp. 1315-1322.
15. Radiological Protection of Patients in Diagnostic and Interventional Radiology, Nuclear Medicine and Radiotherapy. Proceeding of Intern. Conf. Malaga, Spain, 26-30 March Vienna, IAEA, 2001, 587 p.
16. Harbron R.W. [et al.] Patient radiation doses in pediatric interventional cardiology procedures: a review. J. Radiol. Prot, , Vol. 36, No 4, pp. R131-R144.
Review
For citations:
Sarycheva S.S. Estimation of effective dose for children in interventional cardiology. Radiatsionnaya Gygiena = Radiation Hygiene. 2017;10(2):16-22. (In Russ.) https://doi.org/10.21514/1998-426X-2017-10-2-16-22