Effective radiation doses of patients in computed tomography of combined body areas according to phantom dosimetry
https://doi.org/10.21514/1998-426X-2026-19-1-44-55
Abstract
The aim of the study was to perform phantom dosimetry of organs and tissues to assess effective radiation doses to patients during computed tomography of combined areas of the body and calculated based on the readings on the scanner console.
Materials and Methods: The study was performed on СT-scanner Toshiba Aquilion 64 an anthropomorphic tissue-equivalent phantom weighing 73.5 kg using thermoluminescent detectors located in the locations of organs and tissues, under standard computed tomography protocols of two combined body regions: 1 – organs of thorax, abdomen and pelvis; 2 – organs of abdomen and pelvis. The effective radiation dose was calculated using tissue weighting factors according to NRB-99/2009 and ICRP Publication 103 recommendations.
Results and Discussion: The largest contribution to the total radiation dose was found for the gonads, red bone marrow, colon, lungs and stomach (71.8-76.2% in total). It was established that the effective radiation doses to patients during computed tomography, calculated based on the results of phantom dosimetry with TLD detectors, are higher than the values obtained from DLP data on the console of a computer scanner using tissue weighting factors according to ICRP Publication 103 for CT-1 by 1.12 times, for CT-2 by 1.13 times, according to NRB-99/2009 for CT-1 by 1.30 times, for CT-2 by 1.38 times.
Conclusion: Dosimetry on an anthropomorphic tissue-equivalent phantom demonstrates higher patient radiation doses during computed tomography compared to dose values calculated by DLP on the computed tomography scanner console. This demonstrates the importance of periodically comparing patient doses calculated by DLP on the computed tomography scanner console with doses recorded by direct dosimetry under real computed tomography scanning conditions in radiology departments and centers.
Keywords
About the Authors
E. I. MatkevichRussian Federation
Elena I. Matkevich – Candidate of Medical Sciences, Radiologist, Head of the Department of MRI Diagnostics.
23, Marshal Novikov str., Moscow, 123098
P. P. Gantsovsky
Russian Federation
Pavel P. Gantsovsky – Research Associate at the Laboratory of Radiation and Hygienic Research of the A.I. Burnazyan Federal Medical and Biological Center of Federal Medical and Biological Agencyэ
Moscow
A. G. Sivenkov
Russian Federation
Alexander G. Sivenkov – Engineer of Laboratory No. 25 (Laboratory of Radiation and Hygienic research) of the Department of Ionizing and non-ionizing Radiation.
Moscow
A. Yu. Komaro
Russian Federation
Artem Yu. Komarov – Researcher at the Laboratory of Radiation and Hygienic Research at the A.I. Burnazyan Federal Medical and Biological Center of the Federal Medical and Biological Agency.
Moscow
I. L. Yurina
Russian Federation
Irina L. Yurina – Engineer at the Laboratory of Radiation and Hygienic Research at the A.I. Burnazyan Federal Medical and Biological Center of Federal Medical and Biological Agency.
Moscow
A. G. Tsovyanov
Russian Federation
Alexander G. Tsovyanov Candidate of Biological Sciences, Head of the Laboratory of Radiation and Hygienic Research, A.I. Burnazyan Federal Medical and Biological Center of Federal Medical and Biological Agency.
Moscow
A. N. Bashkov
Russian Federation
Andrey N. Bashkov – Candidate of Medical Sciences, Radiologist, Head of the Radiology Center based.
Moscow
References
1. Valentin J. International Commission on Radiation Protection. Managing patient dose in multi-detector computed tomography (MDCT). ICRP Publication 102. Annals of the ICRP. 2007;37(1): 1-79. DOI: 10.1016/j.icrp.2007.09.001.
2. Report to the General Assembly, Sources, Effects and Risks of Ionizing Radiation. United Nations Scientific Committee on the Effects of Atomic Radiation, UNSCEAR 2020/2021. United Nations, New York; 2022. https://www.unscear.org/unscear/publications/2020_2021_1.html (Accessed: 17.03.2026).
3. Balonov MI, Golikov VYu, Vodovatov AV, Chipiga LA, Zvonova IA, Kalnitsky SA, et al. Scientific foundations of radiation protection in modern medicine. Diagnostic radiology: Monograph, manuals. Vol.1. St. Petersburg; 2019. 320 р. (In Russian).
4. Druzhinina PS. Patients` cumulative doses during CTexaminations in Russian clinical practice. Radiatsionnaya Gygiena = Radiation Hygiene. 2025;18(2): 131-140. DOI:10.21514/1998-426X-2025-18-2-131-140. (In Russian).
5. International Commission on Radiation Units and Measurements. ICRU Report No. 87: Radiation dose and image-quality assessment in computed tomography. Journal International Commission on Radiation Units and Measurements. 2012;12(1): 1-149. DOI: 10.1093/jicru/ndt007. DOI:10.1093/jicru/ndt007.
6. Lee C, Won T, Yeom YS, Griffin K, Lee Ch, Kim KP. Organ dose conversion coefficients in CT scans for Korean adult males and females. Nuclear Engineering and Technology. 2022;54(2): 681-688. DOI: 10.1016/j.net.2021.08.008.
7. Cakmak ED, Tuncel N, Sindir B. Assessment of Organ Dose by Direct and Indirect Measurements for a Wide Bore X-Ray Computed Tomography Unit That Used in Radiotherapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology. 2015;4: 132-142. DOI:10.4236/ijmpcero.2015.42017.
8. Lawson M, Berk K, Badawy M, Yujin Qi, Kuganesan Ah, Metcalfe P. Comparison of organ and effective dose estimations from different Monte Carlo simulation-based software methods in infant CT and comparison with direct phantom measurements. Journal of Applied Clinical Medical Physics. 2022;23(6): e13625. DOI: 10.1002/acm2.13625.
9. Zewde N, Ria F, Rehani MM. Organ doses and cancer risk assessment in patients exposed to high doses from recurrent CT exams. European Journal of Radiology. 2022;149: 110224. DOI: 10.1016/j.ejrad.2022.110224.
10. Gao Y, Mahmood U, Liu T, Quinn B, Gollub MJ, Xu XG, et al. Patient-Specific Organ and Effective Dose Estimates in Adult Oncologic CT. American journal of roentgenology. 2020;214(4): 738 – 746. DOI: 10.2214/AJR.19.21197.
11. Matkevich EI, Sivenkov AG, Samoylov AS. The Morbidity Structure of Personnel Working with Man-Made Radiation Sources Referred for Computed Tomography at the State Research Center — Burnasyan Federal Medical Biophysical Center of Federal Medical Biological Agency in 2020–2024. Radiatsionnaya biologiya. Radioekologiya = Radiation biology. Radioecology. 2025;65(6): 597-613. DOI: 10.7868/S3034590125060031. (In Russian).
12. The Alderson Radiation Therapy Phantom. Available from: https://rsdphantoms.com/product/the-alderson-radiationtherapy-phantom/ (Accessed: 17.03.2026).
13. Thermoluminescent Dosimetric Units «DVG-02TM». Description of the Measuring Instrument Type. Appendix to Certificate No. 45087 on Approval of the Measuring Instrument Type. Federal State Unitary Enterprise Research Institute of Industrial and Marine Medicine (NIIPMM). St. Petersburg: 2011. https://fgis.gost.ru/fundmetrology/cm/mits/497e715c7627-eb6c-396a-dd2871f03517 (Accessed: 17.03.2026). (In Russian).
14. Universal dosimeter DKS-101. Operation manual. GKPS13.00.00.000RE. Moscow, 2009. http://медтехатом.рф/upload/iblock/1e0/1e0571199faeb4116640a1fbf1666162.pdf (Accessed: 17.03.2026). (In Russian).
15. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Annals of the ICRP. 2007; 37 (2-4): 1-332. DOI: 10.1016/j.icrp.2007.10.003. (In Russian).
16. Druzhinina YuV, Lantukh ZA, Vodovatov AV, Morozov SP, Ryzhov SA, Sokolov EN, et al. Development and Application of Diagnostic Reference Levels for Adult Patients in Diagnostic Radiology. Meditsinskaya Fizika = Medical Physics. 2022;93(1): 81-96. DOI: 10.52775/1810-200X-2022-93-1-81-96. (In Russian).
17. Chipiga L, Golikov V, Shleenkova E, Pozdnyakov A. Estimation of the conversion coefficients from dose length product to effective dose from whole body computed tomography examination using anthropomorphic phantoms. Meditsinskaya Fizika = Medical Physics. 2016;72(4): 55-62. (In Russian).
18. Druzhinina PS, Chipiga LA, Vodovatov AV, Soldatov IV, Lantukh ZA, Tolkachev KV. Determination of the conversion coefficients from the dose-length product to the effective dose for CT-examinations of the whole body including lower extremitie. Radiatsionnaya Gygiena = Radiation Hygiene. 2024;17(4): 126-134. DOI:10.21514/1998-426X-2024-17-4-126-134. (In Russian).
19. Brenner DJ. It is time to retire the computed tomography dose index (CTDI) for CT quality assurance and dose optimization. For the proposition. Meditsinskaya Fizika = Medical Physics. 2006;33(5): 1189-90. DOI: 10.1118/1.2173933.
20. Parsi M, Sohrabi M, Mianji F, Paydar R. A «Quality-controlbased correction method» for displayed dose indices on CT scanner consoles in patient dose surveys. Physica Medica. 2017;38: 88-92. DOI: 10.1016/j.ejmp.2017.05.054.
Review
For citations:
Matkevich E.I., Gantsovsky P.P., Sivenkov A.G., Komaro A.Yu., Yurina I.L., Tsovyanov A.G., Bashkov A.N. Effective radiation doses of patients in computed tomography of combined body areas according to phantom dosimetry. Radiatsionnaya Gygiena = Radiation Hygiene. 2026;19(1):44-55. (In Russ.) https://doi.org/10.21514/1998-426X-2026-19-1-44-55
JATS XML