Radiation doses for patients undergoing digital mammography in Saint Petersburg and Leningrad Region
https://doi.org/10.21514/1998-426X-2026-19-2-34-42
Abstract
The gold standard for early diagnosis of malignant breast neoplasms is mammography (X-ray examination of the breast) performed using digital and analogue X-ray units. This procedure is routinely conducted for all women aged 40 to 75 years on a regular basis (once every one or two years); mammography of both breasts is performed in two projections (four X-ray images). Doses from mammography, in accordance with the Norms of the Radiation Safety NRB 99/2009, shall not exceed 1 mSv per year. Hence, reliable assessment of patient doses for this procedure is important.
The aim of this study was to evaluate the patient doses from mammographic procedures in medical facilities in St. Petersburg and the Leningrad Region.
Materials and Methods: Data collection was performed at 99 medical facilities in St. Petersburg and the Leningrad Region for 141 digital mammography units during the period 2023–2025. For each mammography unit, mean glandular dose and effective dose was assessed for various compressed breast thicknesses (2, 4, 6, 8, and 10 cm) and projections (craniocaudal and mediolateral oblique), in accordance with the methodological guidelines MU 2.6.1.2944-11.
Results and Discussion: The mean effective doses per single mammographic image in either the craniocaudal or mediolateral oblique projection varied from 0.05 to 0.07 mSv; effective doses per complete mammographic examination (two craniocaudal and two mediolateral oblique images) varied from 0.2 to 0.3 mSv. At the same time, substantial differences in patient dose ranges were observed: the ratio between the minimum and maximum mean glandular dose and effective dose reached up by the factor of 60, respectively. The primary factors determining patient exposure doses were the exposure and the X-ray tube output. No significant correlation was found between patient doses and tube voltage, focal-image distance, or the anode/filter combination. Compressed breast thickness also did not significantly influence the effective dose.
Conclusion: Although the mean patient exposure levels during digital mammography are acceptable and comparable with international practice, patient doses for several mammography units are anomalously high. For such units, the implementation of patient radiation protection optimization measures is a priority.
Keywords
About the Authors
A. V. VodovatovRussian Federation
Aleksandr V. Vodovatov, Candidate of Biological Sciences, Head of Laboratory
Laboratory of Radiation Hygiene of Medical Facilities
197101; 8, Mira Str.; Saint Petersburg
РИНЦ Author ID: 859255
E. A. Kosarlukova
Russian Federation
Elena A. Kosarlukova, Acting Junior Researcher
Information Analytical Center
Saint Petersburg
K. E. Kostina
Russian Federation
Ksenia E. Kostina, Acting Junior Researcher
Laboratory of Radiation Control
Saint Petersburg
L V. Repin
Russian Federation
Leonid V. Repin, Junior Researcher
Information Analytical Center
Saint Petersburg
РИНЦ Author ID: 653792
G. A. Gorsky
Russian Federation
Grigory A. Gorsky, Candidate of Medical Sciences, Deputy Director of the Innovation Work, Docent
Saint Petersburg
РИНЦ Author ID: 656343
D. A. Likhun
Russian Federation
Dmitry A. Likhun, Head of the Department
Radiation Hygiene Supervision Department
Saint Petersbur
References
1. Cancer today. GLOBOCAN. Available from: https://gco.iarc.fr [Accessed 06 April 2026].
2. Drapkina OM, Kaprin AD, Almazova II, Drozdova LYu, Ivanova ES, Rozhkova NI, et al. Screening of malignant neoplasms of the breast as part of the medical examination of certain groups of the adult population. Guidelines. Pervichnaya mediko-sanitarnaya pomoshch = Primary Health Care (Russian Federation). 2024;1(1): 63-80. DOI: 10.15829/3034-4123-2024-5. (In Russian).
3. Breast cancer. World Health Organization. Available from: https://www.who.int/ru [Accessed 06 April 2026]. (In Russian).
4. Kaprin AD, Starinsky VV, Shakhzadova AO. Malignant neoplasms in Russia in 2023 (morbidity and mortality). Moscow: P. Hertsen Moscow Oncology Research Institute - a branch of National Medical Research Radiological Centre of the Ministry of Health of the Russian Federation; 2024. 276 p.
5. Bontrager KL, Lampignano JP. Textbook of radiographic positioning and related anatomy. 8<sup>th</sup> ed. Mosby; 2014. 826 p.
6. Vodovatov AV. Improvement of radiation safety standards. Part 1. Appropriatennes of the limitation of the medical exposure of healthy individuals. Radiatsionnaya Gygiena = Radiation Hygiene. 2018;11(3): 115-124. DOI: 10.21514/1998-426X-2018-11-3-115-124. (In Russian).
7. Hendrick RE. Radiation doses and cancer risks from breast imaging studies. Radiology. 2010;257(1): 246-253.
8. Hendrick RE, Pisano ED, Averbukh A, Moran C, Berns EA, Yaffe MJ, et al. Comparison of acquisition parameters and breast dose in digital mammography and screen-film mammography in the American College of Radiology Imaging Network digital mammographic imaging screening trial. AJR. American Journal of Roentgenology. 2010;194(2): 362-369. DOI: 10.2214/AJR.08.2114.
9. Sulieman A, Serhan O, Al-Mohammed HI, Mahmoud MZ, Alkhorayef M, Alonazi B, et al. Estimation of cancer risks during mammography procedure in Saudi Arabia. Saudi Journal of Biological Sciences. 2019;26(6): 1107-1111. DOI: 10.1016/j.sjbs.2018.10.005.
10. Baek JE, Kang BJ, Kim SH, Lee HS. Radiation dose affected by mammographic composition and breast size: first application of a radiation dose management system for full-field digital mammography in Korean women. World Journal of Surgical Oncology. 2017;15(1): 38. DOI: 10.1186/s12957-017-1107-6.
11. Amiri F, Jacob B, Dkhar W, Mustaffa SP, Kadavigere R, Pradhan A, et al. Evaluation of mean glandular dose in relation to breast density in digital mammography. Egyptian Journal of Radiology and Nuclear Medicine. 2025;56: 161. DOI: 10.1186/s43055-025-01586-y.
12. Bečirović H, Kunosić S, Halilović M, Lasić I, Galić S, Bajrović J. Radiation Exposure of Patients in Mammography : An Overview of the 15-Year Practice. In: Badnjevic A, Gurbeta Pokvić L. (eds). CMBEBIH 2021 : Proceedings of the International Conference on Medical and Biological Engineering, CMBEBIH 2021, 21-24 April 2021, Mostar, Bosnia and Herzegovina. Cham: Springer; 2021. Vol. 84. P. 809-816. DOI: 10.1007/978-3-030-73909-6_92.
13. Warren LM, Dance DR, Young KC. Radiation risk of breast screening in England with digital mammography. The British Journal of Radiology. 2016;89(1067): 20150897. DOI: 10.1259/bjr.20150897.
14. Du X, Yu N, Zhang Y, Wang JT. The relationship of the mean glandular dose with compressed breast thickness in mammography. Journal of Public Health and Emergency. 2017;1: 32 DOI: 10.21037/jphe.2017.03.10.
15. Bor D, Tukel S, Olgar T, Toklu T, Aydin E, Akyol O. Investigation of mean glandular dose versus compressed breast thickness relationship for mammography. Radiation Potection Dosimetry. 2008;129(1-3): 160-164. DOI: 10.1093/rpd/ncn140.
16. Lalji UC, Jeukens CR, Houben I, Nelemans PJ, van Engen RE, van Wylick E, et al. Evaluation of low-energy contrast-enhanced spectral mammography images by comparing them to full-field digital mammography using EUREF image quality criteria. European Radiology. 2015;25(10): 2813-2820. DOI: 10.1007/s00330-015-3695-2.
17. Sechopoulos I, Suryanarayanan S, Vedantham S, D'Orsi CJ, Karellas A. Radiation dose to organs and tissues from mammography: Monte Carlo and phantom study. Radiology. 2008;246(2): 434-443. DOI: 10.1148/radiol.2462070256.
Review
For citations:
Vodovatov A.V., Kosarlukova E.A., Kostina K.E., Repin L.V., Gorsky G.A., Likhun D.A. Radiation doses for patients undergoing digital mammography in Saint Petersburg and Leningrad Region. Radiatsionnaya Gygiena = Radiation Hygiene. 2026;19(2):34-42. (In Russ.) https://doi.org/10.21514/1998-426X-2026-19-2-34-42
JATS XML